answersLogoWhite

0

When switching a microscope from low power to high power, you should first ensure that the objective lens is clean and in the correct position. Next, carefully rotate the nosepiece to switch to the high-power objective lens. It is important to avoid touching the lens with your fingers and to refocus the image using the fine adjustment knob to prevent damage to the lens and obtain a clear image at the higher magnification.

User Avatar

ProfBot

5mo ago

What else can I help you with?

Continue Learning about Biology

What two things must be true before the microscope can be switched to high power to zoom in on a specimen?

Before switching to high power, the specimen must be in focus and positioned in the center of the field of view under low power magnification. Additionally, the light source should be properly adjusted to provide clear illumination of the specimen at high power.


What part of a microscope rotates to change from the low power objective lens to the high power objective lens?

The revolving nosepiece or turret rotates to change from the low power objective lens to the high power objective lens on a microscope. This component allows for easy switching between different objective lenses without the need to manually remove and insert them.


What is the usual magnifying power of the high power objective on your compound microscope?

The high power objective on a compound microscope typically has a magnifying power of 40x or 50x.


What is a high power objective in a microscope?

A high power objective in a microscope is a lens that provides high magnification of the specimen being viewed. It typically magnifies the image by 40x to 100x, allowing for detailed observation of fine structures. High power objectives are used for examining smaller details and are typically used after lower magnification objectives.


Which part of a microscope should be used with the low power objective but not with the high power objective?

The coarse focus knob should be used with the low power objective but not with the high power objective. The point of using the coarse focus knob is to bring the specimen into general focus before fine-tuning with the high power objective, which requires more precise adjustments.

Related Questions

When switching from the low-power objective to the high-power objective should you open or close the diaphragm Explain your answer?

When working with a microscope you should always take procauctions but one question I am asked a lot is When working with a microscope do you open or close the diaphram when working with the medium- and high-power objective nobs? The answer is you keep it open if you don't you won't be abler to see any thing.


Why should you never lower the stage in a microscope before hitting high power?

You should never lower the stage in a microscope before switching to high power because doing so can cause the objective lens to come into contact with the slide, potentially damaging both the lens and the slide. This can lead to scratches on the lens or breakage of the slide, compromising your samples. Additionally, maintaining a proper working distance helps ensure clear and focused images at higher magnifications. Always adjust the focus from low power before switching to high power.


Why must the specimen be centered before switching to high power on a microscope?

Centering the specimen before switching to high power on a microscope helps ensure that the area you want to observe is in focus and at the center of the field of view. This is important because high power magnification has a narrower field of view and lower depth of field, making it more challenging to locate and focus on the specimen if it is not first centered.


How would you have to adjust the diaphragm on a microscope after switching form low to high power?

After switching the microscope from low to high power, you may need to readjust the diaphragm to regulate the amount of light entering the lens. This adjustment helps maintain optimal image brightness and clarity at different magnifications. Start by opening the diaphragm fully and then gradually close it until the image is clear and properly illuminated at high power.


When switching from low power objective to high power objective should you open or close diaphragm?

close


When switching from low-power objective to high-power objective should you open or close to diaphragm?

close


Does the switch from low power to high power change the position of the image under microscope?

No, switching from low power to high power magnifies the image but does not change its position. The position of the image remains the same as long as the slide or specimen is not moved.


What two things must be true before the microscope can be switched to high power to zoom in on a specimen?

Before switching to high power, the specimen must be in focus and positioned in the center of the field of view under low power magnification. Additionally, the light source should be properly adjusted to provide clear illumination of the specimen at high power.


How dows the procedure for using the microscope differ under high power as opposed to low power?

When using a microscope under high power, you typically start by locating the specimen using the coarse adjustment knob on low power before switching to high power for a more detailed view. Under high power, you may need to use the fine adjustment knob more delicately to focus due to the increased magnification. Additionally, the field of view is smaller under high power compared to low power.


What part of a microscope rotates to change from the low power objective lens to the high power objective lens?

The revolving nosepiece or turret rotates to change from the low power objective lens to the high power objective lens on a microscope. This component allows for easy switching between different objective lenses without the need to manually remove and insert them.


What should you remember when handling microscope slide?

Do not allow the high power objective to tough the lenses


What is the usual magnifying power of the high power objective on your compound microscope?

The high power objective on a compound microscope typically has a magnifying power of 40x or 50x.