answersLogoWhite

0

What else can I help you with?

Continue Learning about Natural Sciences

Why is it important to focus specimens under the Low Power Objectve before observing it under the High Power Objective?

It allows you to center the specimen, and if it is set up correctly, then when you switch objectives it should still be there and only need fine focusing. It also reduces the chance of crushing the specimen and/or breaking the slide or the lens as the LPO is further away from the slide than the HPO.


Say you were observing a specimen on the low power objective when you switch to high power the specimen is no longer visible why?

The high power objective has a shorter focal length and narrower field of view compared to the low power objective. This can cause the specimen to move out of the field of view when switching from low to high power, especially if it was not centered properly. Adjusting the mechanical stage to re-center the specimen or refocusing the microscope may help bring the specimen back into view on high power.


What would happen to the center of the slide as the magnification increased?

As magnification increases, the center of the slide would appear larger and more detailed. This is because magnification enlarges the image of the specimen on the slide, allowing you to see finer features with greater clarity at the center.


Why is it important to start with a low magnification when using the microscope?

So you can determine the actual size of what you are looking at.


Why do you need to center your specimen under the lowest objective lens before switching to a higher objective lens?

Centering your specimen under the lowest objective lens ensures that the area of interest is in the field of view when you switch to a higher magnification. This minimizes the risk of losing the target area and allows for a smoother transition between objectives. Additionally, it helps maintain proper focus and avoids potential damage to the slide or the lens when adjusting for higher magnifications.

Related Questions

What happens if you did not center the specimen before moving to higher?

If you do not center the specimen before moving to higher magnifications on a microscope, the specimen may move out of the field of view or become difficult to find at the higher magnification. This can lead to difficulties in focusing and observing the specimen accurately. It's important to center the specimen at lower magnifications before moving to higher ones to maintain a clear view.


Why must the specimen be centered before switching to high power on a microscope?

Centering the specimen before switching to high power on a microscope helps ensure that the area you want to observe is in focus and at the center of the field of view. This is important because high power magnification has a narrower field of view and lower depth of field, making it more challenging to locate and focus on the specimen if it is not first centered.


Why do you need to focus specimen under LPO first?

Focusing a specimen under low power objective (LPO) allows you to find and center the specimen on the slide before moving to higher magnifications. This ensures that the sample is in the field of view and helps in adjusting the focus and moving to higher magnifications without losing sight of the specimen.


Why is it important to focus specimens under the Low Power Objectve before observing it under the High Power Objective?

It allows you to center the specimen, and if it is set up correctly, then when you switch objectives it should still be there and only need fine focusing. It also reduces the chance of crushing the specimen and/or breaking the slide or the lens as the LPO is further away from the slide than the HPO.


What two things must be true before the microscope can be switched to high power to zoom in on a specimen?

Before switching to high power, the specimen must be in focus and positioned in the center of the field of view under low power magnification. Additionally, the light source should be properly adjusted to provide clear illumination of the specimen at high power.


Why must you always use the low power lens when you are first focusing on a specimen?

It is supposed to help you center on the specimen.


Lens used to locate specimen on a microscope?

The lens used to locate a specimen on a microscope is typically the lowest magnification lens, known as the scanning lens or low power objective. This lens provides a wide field of view, making it easier to find and center the specimen before switching to higher magnification lenses for detailed viewing.


Why must the specimen be centered before switching from low power to high power?

Because high power takes up a much smaller area, and you may not be able to find the specimen if it's off to one side. C:


Why should you center the image on a microscope before switching to a higher power?

Centering the image on a microscope before switching to a higher power ensures that the object of interest stays in the field of view. This helps maintain focus and prevents the risk of losing sight of the specimen when magnifying it further.


What important center was Babylon located in?

Important center of Trade


Is the great center giveaway a scam?

There is no definitive evidence to suggest that the Great Center Giveaway is a scam. It is important to research and verify the legitimacy of any giveaway before participating.


Say you were observing a specimen on the low power objective when you switch to high power the specimen is no longer visible why?

The high power objective has a shorter focal length and narrower field of view compared to the low power objective. This can cause the specimen to move out of the field of view when switching from low to high power, especially if it was not centered properly. Adjusting the mechanical stage to re-center the specimen or refocusing the microscope may help bring the specimen back into view on high power.