the same way as under low-power on a microscope: carefully
The microscope lens located nearest the specimen is called the objective lens. It is responsible for gathering light from the specimen and magnifying the image, allowing for detailed observation. Different objective lenses can provide varying levels of magnification, typically ranging from low to high power.
A high power objective lens is a microscope lens with a high magnification level, used for viewing specimens in fine detail. It allows for closer inspection of specific features or structures of the specimen.
When focusing on a specimen under high power, you should use the fine adjustment knob. The coarse adjustment knob is typically used only under low power to prevent damage to the slide or objective lens. The fine adjustment allows for precise focusing, which is crucial at high magnifications to achieve a clear and sharp image of the specimen. Always ensure that you start with the lowest power objective before switching to high power for safety.
I believe it's the coarse adjustment.
Fine Adjustment knob
It means you have a good microscope.
Dropping the microscope or mishandling it can lead to damage, as well as using excessive force when adjusting the focus or moving the slides. Using improper cleaning methods or harsh chemicals can also damage the microscope and specimen.
The magnification of the specimen under low power optics, lpo, is 10 times and the magnification of the specimen under high power optics, hpo, depends on the power of the microscope but is usually at least 500 times or more.
The specimen appeared as a collection of individual cells or particles, which were visible at a very high magnification. The color, shape, and arrangement of the specimen's components could be observed in detail under the microscope.
Before switching to high power, the specimen must be in focus and positioned in the center of the field of view under low power magnification. Additionally, the light source should be properly adjusted to provide clear illumination of the specimen at high power.
You use the low power lens to find the specimen with a microscope. Once you have found the specimen on low power, you can switch to high power for greater magnification. Just be sure to use only the fine adjustment knobs for focusing on high power so the high power objective won't hit the slide.
When using a microscope under high power, you typically start by locating the specimen using the coarse adjustment knob on low power before switching to high power for a more detailed view. Under high power, you may need to use the fine adjustment knob more delicately to focus due to the increased magnification. Additionally, the field of view is smaller under high power compared to low power.
If the slide is not moved and the high-power objective lens is switched into place, the specimen will appear larger and more detailed, but it may also be out of focus. The high-power lens provides a narrower field of view, so only a portion of the specimen will be visible. It is essential to refocus the microscope to get a clear image of the specimen under high power.
No, switching from low power to high power magnifies the image but does not change its position. The position of the image remains the same as long as the slide or specimen is not moved.
Course Focus Knob- low Fine Focus Knob- high -MC
first view with low power then gradually increase the power. hope it helped you!
Centering the specimen before switching to high power on a microscope helps ensure that the area you want to observe is in focus and at the center of the field of view. This is important because high power magnification has a narrower field of view and lower depth of field, making it more challenging to locate and focus on the specimen if it is not first centered.