answersLogoWhite

0

What else can I help you with?

Continue Learning about Natural Sciences

Say you were observing a specimen on the low power objective when you switch to high power the specimen is no longer visible why?

The high power objective has a shorter focal length and narrower field of view compared to the low power objective. This can cause the specimen to move out of the field of view when switching from low to high power, especially if it was not centered properly. Adjusting the mechanical stage to re-center the specimen or refocusing the microscope may help bring the specimen back into view on high power.


Why is it important to center a specimen on low power before attemping to focus on it at high power?

Centering a specimen on low power ensures that it is within the field of view and makes it easier to locate and focus on at high power. If a specimen is not centered on low power, it may be difficult to find and focus on when switching to high power, leading to potential damage to the specimen or microscope lenses. Additionally, centering the specimen helps to achieve the best image quality and clarity at high magnification.


The specimen must be centered in the field of view on low power before going to high power?

Because when looking at it from far, then you can see more of the specimen, However, when you're looking at it closer up in a high power mode, then you will see less of the object, but much closer.


Does the switch from low power to high power change the position of the image under microscope?

No, switching from low power to high power magnifies the image but does not change its position. The position of the image remains the same as long as the slide or specimen is not moved.


Why is it important to focus specimens under the Low Power Objectve before observing it under the High Power Objective?

It allows you to center the specimen, and if it is set up correctly, then when you switch objectives it should still be there and only need fine focusing. It also reduces the chance of crushing the specimen and/or breaking the slide or the lens as the LPO is further away from the slide than the HPO.

Related Questions

Why must the specimen be centered before switching to high power on a microscope?

Centering the specimen before switching to high power on a microscope helps ensure that the area you want to observe is in focus and at the center of the field of view. This is important because high power magnification has a narrower field of view and lower depth of field, making it more challenging to locate and focus on the specimen if it is not first centered.


Why must the specimen be centered before switching from low power to high power?

Because high power takes up a much smaller area, and you may not be able to find the specimen if it's off to one side. C:


Say you were observing a specimen on the low power objective when you switch to high power the specimen is no longer visible why?

The high power objective has a shorter focal length and narrower field of view compared to the low power objective. This can cause the specimen to move out of the field of view when switching from low to high power, especially if it was not centered properly. Adjusting the mechanical stage to re-center the specimen or refocusing the microscope may help bring the specimen back into view on high power.


What two things must be true before the microscope can be switched to high power to zoom in on a specimen?

Before switching to high power, the specimen must be in focus and positioned in the center of the field of view under low power magnification. Additionally, the light source should be properly adjusted to provide clear illumination of the specimen at high power.


Why is it important to center a specimen on low power before attemping to focus on it at high power?

Centering a specimen on low power ensures that it is within the field of view and makes it easier to locate and focus on at high power. If a specimen is not centered on low power, it may be difficult to find and focus on when switching to high power, leading to potential damage to the specimen or microscope lenses. Additionally, centering the specimen helps to achieve the best image quality and clarity at high magnification.


Why must the specimen be centered before switching to highpower in a microscope?

Centering the specimen before switching to high power helps ensure that you are viewing the area of interest in focus. It helps prevent the objective lens from hitting the slide, which can damage both the lens and the specimen. Additionally, centering the specimen can improve the image quality by reducing glare and shadowing.


The specimen must be centered in the field of view on low power before going to high power?

Because when looking at it from far, then you can see more of the specimen, However, when you're looking at it closer up in a high power mode, then you will see less of the object, but much closer.


Does the switch from low power to high power change the position of the image under microscope?

No, switching from low power to high power magnifies the image but does not change its position. The position of the image remains the same as long as the slide or specimen is not moved.


Why the specimen be centered in field of view on low power before going to high power?

Because it is too hard to find a specimen on high power. Easier to find them and focus them on low power and then increase the magnification and fine-tune the focus to get a better, closer look.


What is the lens that is used to find the specimen on a microscope?

You use the low power lens to find the specimen with a microscope. Once you have found the specimen on low power, you can switch to high power for greater magnification. Just be sure to use only the fine adjustment knobs for focusing on high power so the high power objective won't hit the slide.


Why is it important to focus specimens under the Low Power Objectve before observing it under the High Power Objective?

It allows you to center the specimen, and if it is set up correctly, then when you switch objectives it should still be there and only need fine focusing. It also reduces the chance of crushing the specimen and/or breaking the slide or the lens as the LPO is further away from the slide than the HPO.


How dows the procedure for using the microscope differ under high power as opposed to low power?

When using a microscope under high power, you typically start by locating the specimen using the coarse adjustment knob on low power before switching to high power for a more detailed view. Under high power, you may need to use the fine adjustment knob more delicately to focus due to the increased magnification. Additionally, the field of view is smaller under high power compared to low power.