answersLogoWhite

0

Chromatography

Chromatography is a set of techniques used to separate mixtures in a chemistry lab setting. The word chromatography means "color writing" and some of the first experiments in chromatography involved separating chlorophyll in plants.

649 Questions

How do you solve the Rf value in Column Chromatography?

Measure the distance from your marked pencil line to the solvent front. Also measure the distance fomr the marked pencil line to the middle of the spot, and divide this by the first distance. This value will be between 0 and 1.

Calibration of gas chromatography?

What is the precedure for calibration of gas chromatography? How calibration is perform for gas chromatography? What is the precedure for calibration of gas chromatography? How calibration is perform for gas chromatography?

What is void volume in hplc?

The void volume in HPLC is the volume of the column that is not occupied by the stationary phase. It represents the space where mobile phase flows through without interacting with the stationary phase or sample components. A large void volume can lead to poor resolution of peaks in chromatography.

What is definition of area normalisation method in HPLC?

The goal of area normalization is to correct for sample size discrepances, that in a negative way affects the sum of all measured solutes.

Example

If 4 peaks are being measured and the sum total area of all peaks turns out

to be less than or greater than 100 percent, Normalization corrects for this

and shifts all peaks by the needed percentage to bring them to a sum of 100

percent. If the sum total equals 90 percent, then normalization will shift up

each individual peak 10 percent, which will bring the sum up to 100 percent..

Where are carotenoids seen with chromatography?

An analytical method for the profiling and quantitative determination of carotenoids in bacteria is described. Exhaustive extraction of the pigments from four selected bacterial strains required treatment of the cells with potassium hydroxide or liquefied phenol or both before the addition of the extracting solvent (methanol or diethyl ether). The carotenoids in the extracts were separated by nonaqueous reversed-phase liquid chromatography in conjunction with photodiode array absorption detection. The identity of a peak was considered definitive only when both its retention time and absorption spectrum, before and after chemical reactions, matched those of a reference component. In the absence of the latter, most peaks could be tentatively identified. Two examples illustrate how in the analysis of pigmented bacteria errors may result from using nonchromatographic procedures or liquid chromatographic methods lacking sufficient criteria for peak identification. Carotenoids of interest were determined quantitatively when the authentic reference substance was available or, alternatively, were determined semiquantitatively http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=203224

What does Rf Values in chromatography represent?

Rf values represent the ratio between the distance a component travels and the distance the solvent moves in thin-layer chromatography or paper chromatography. It is a dimensionless quantity used to identify and compare compounds based on their ability to move with the solvent.

Who uses chromatography?

Chromatography is used in pharmaceutical industries for the synthesis of various drugs to check whether the product is synthesized from the reactants or not. It is used in petrochemical industries to check the purity of the fuel and its additive. It is also used by forensic scientists & helps them in solving different cases.

What is the use of gas chromatography in food industry?

Chromatographydates to 1903 in the work of the Russian scientist, Mikhail Semenovich Tswett. Germangraduate student Fritz Priordeveloped solid state gas chromatography in 1947. Archer John Porter Martin, who was awarded the Nobel Prize for his work in developing liquid-liquid (1941) and paper (1944) chromatography, laid the foundation for the development of gas chromatography and he later produced liquid-gas chromatography (1950). Erika Cremerlaid the groundwork, and oversaw much of Prior's work.

[edit]GC analysis

A gas chromatograph is a chemical analysis instrument for separating chemicals in a complex sample. A gas chromatograph uses a flow-through narrow tube known as the column, through which different chemical constituents of a sample pass in a gas stream (carrier gas, mobile phase) at different rates depending on their various chemical and physical properties and their interaction with a specific column filling, called the stationary phase. As the chemicals exit the end of the column, they are detected and identified electronically. The function of the stationary phase in the column is to separate different components, causing each one to exit the column at a different time (retention time). Other parameters that can be used to alter the order or time of retention are the carrier gas flow rate, column length and the temperature.

In a GC analysis, a known volume of gaseous or liquid analyte is injected into the "entrance" (head) of the column, usually using a microsyringe(or, solid phase microextraction fibers, or a gas source switching system). As the carrier gas sweeps the analyte molecules through the column, this motion is inhibited by the adsorption of the analyte moleculeseither onto the column walls or onto packing materials in the column. The rate at which the molecules progress along the column depends on the strength of adsorption, which in turn depends on the type of molecule and on the stationary phase materials. Since each type of molecule has a different rate of progression, the various components of the analyte mixture are separated as they progress along the column and reach the end of the column at different times (retention time). A detector is used to monitor the outlet stream from the column; thus, the time at which each component reaches the outlet and the amount of that component can be determined. Generally, substances are identified (qualitatively) by the order in which they emerge (elute) from the column and by the retention time of the analyte in the column.

[edit]Physical components

Diagram of a gas chromatograph.

[edit]Autosamplers

The autosampler provides the means to introduce a sample automatically into the inlets. Manual insertion of the sample is possible but is no longer common. Automatic insertion provides better reproducibility and time-optimization.

Different kinds of autosamplers exist. Autosamplers can be classified in relation to sample capacity (auto-injectors vs. autosamplers, where auto-injectors can work a small number of samples), to robotic technologies (XYZ robot vs. rotating robot - the most common), or to analysis:

Traditionally autosampler manufacturers are different from GC manufacturers and currently no GC manufacturer offers a complete range of autosamplers. Historically, the countries most active in autosampler technology development are the United States, Italy, Switzerland, and the United Kingdom.

[edit]Inlets

The column inlet (or injector) provides the means to introduce a sample into a continuous flow of carrier gas. The inlet is a piece of hardware attached to the column head.

Common inlet types are:

  • S/SL (split/splitless) injector; a sample is introduced into a heated small chamber via a syringe through a septum - the heat facilitates volatilizationof the sample and sample matrix. The carrier gas then either sweeps the entirety (splitless mode) or a portion (split mode) of the sample into the column. In split mode, a part of the sample/carrier gas mixture in the injection chamber is exhausted through the split vent. Split injection is preferred when working with samples with high analyte concentrations (>0.1%) whereas splitless injection is best suited for trace analysis with low amounts of analytes (<0.01%). In splitless mode the split valve opens after a pre-set amount of time to purge heavier elements that would otherwise contaminate the system. This pre-set (splitless) time should be optimized, the shorter time (e.g., 0.2 min) ensures less tailing but loss in response, the longer time (2 min) increases tailing but also signal.
  • On-column inlet; the sample is here introduced directly into the column in its entirety without heat.
  • PTV injector; Temperature-programmed sample introduction was first described by Vogt in 1979.[citation needed] Originally Vogt developed the technique as a method for the introduction of large sample volumes (up to 250 µL) in capillary GC. Vogt introduced the sample into the liner at a controlled injection rate. The temperature of the liner was chosen slightly below the boiling point of the solvent. The low-boiling solvent was continuously evaporated and vented through the split line. Based on this technique, Poy developed the programmed temperature vaporising injector; PTV. By introducing the sample at a low initial liner temperature many of the disadvantages of the classic hot injection techniques could be circumvented.[citation needed]
  • Gas source inlet or gas switching valve; gaseous samples in collection bottles are connected to what is most commonly a six-port switching valve. The carrier gas flow is not interrupted while a sample can be expanded into a previously evacuated sample loop. Upon switching, the contents of the sample loop are inserted into the carrier gas stream.
  • P/T (Purge-and-Trap) system; An inert gas is bubbled through an aqueous sample causing insoluble volatile chemicals to be purged from the matrix. The volatiles are 'trapped' on an absorbent column (known as a trap or concentrator) at ambient temperature. The trap is then heated and the volatiles are directed into the carrier gas stream. Samples requiring preconcentration or purification can be introduced via such a system, usually hooked up to the S/SL port.

The choice of carrier gas (mobile phase) is important, hydrogen has a larger range of flowrates that are comparable to helium in efficiency . However, helium, may be more efficient and provide the best separation if flow rates are optimised. Helium is non-flammable, and works with a greater number of detectors. Therefore, helium is the most common carrier gas used. Historical use rather than rational consideration may contribute to its continued preferential use of helium.

[edit]Detectors

The most commonly used detectors are the flame ionization detector (FID) and the thermal conductivity detector (TCD). Both are sensitive to a wide range of components, and both work over a wide range of concentrations. While TCDs are essentially universal and can be used to detect any component other than the carrier gas (as long as their thermal conductivities are different from that of the carrier gas, at detector temperature), FIDs are sensitive primarily to hydrocarbons, and are more sensitive to them than TCD. However, an FID cannot detect water. Both detectors are also quite robust. Since TCD is non-destructive, it can be operated in-series before an FID (destructive), thus providing complementary detection of the same analytes.[3]

Other detectors are sensitive only to specific types of substances, or work well only in narrower ranges of concentrations. They include:

  • Thermal Conductivity detector (TCD), this common detector relies on the thermal conductivity of matter passing around a tungsten -rhenium filament with a current traveling through it.[4]In this set up helium or nitrogen serve as the carrier gas because of their relatively high thermal conductivity which keep the filament cool and maintain uniform resistivity and electrical efficiency of the filament.[5][4]However, when analyte molecules elute from the column, mixed with carrier gas, the thermal conductivity decreases and this causes a detector response.[5]The response is due to the decreased thermal conductivity causing an increase in filament temperature and resistivity resulting in flucuations in voltage.[4]Detector sensitivity is proportional to filament current while its inversely proportional to the immediate environmental temperature of that dector as well as flow rate of the carrier gas.[4]
  • Flame Ionization detector (FID), in this common detector electrodes are placed adjacent to a flame fueled by hydrogen / air near the exit of the column, and when carbon containing compounds exit the column they are pyrolyzed by the flame.[5][4]This detector works only for organic / hydrocarbon containing compounds due to the ability of the carbons to form cations and electrons upon pyrolysis which generates a current between the electrodes.[5][4]The increase in current is translated and appears as a peak in a chromatogram. FIDs have low detection limits (a few picograms per second, but they are unable to generate ions from carbonyl containing carbons.[4]FID compatible carrier gasses include nitrogen, helium, and argon.[5][4]
  • Catalytic combustion detector (CCD), which measures combustible hydrocarbons and hydrogen.
  • Discharge ionization detector (DID), which uses a high-voltage electric discharge to produce ions.
  • Dry electrolytic conductivity detector (DELCD), which uses an air phase and high temperature (v. Coulsen) to measure chlorinated compounds.
  • Electron capture detector (ECD), which uses a radioactive beta particle (electron) source to measure the degree of electron capture. ECD are used for the detection of molecules containing electronegative / withdrawing elements and functional groups like halogens, carbonyl, nitriles, nitro groups, and organometalics.[5][4]In this type of detector either nitrogen or 5% methane in argon is used as the mobile phase carrier gas.[5][4]The carrier gas passes between two electrodes placed at the end of the column, and adjacent to the anode (negative electrode) resides a radioactive foil such as 63Ni.[5][4]The radioactive foil emits a beta particle (electrode) which collides with and ionizes the carrier gas to generate more ions resulting in a current.[5][4]When analyte molecules with electronegative / withdrawing elements or functional groups electrons are captured which results in a decrease in current generating a detector response.[5][4]
  • Flame photometric detector (FPD),which uses a photomultiplier tube to detect spectral lines of the compounds as they are burned in a flame. Compounds eluting off the coloumn are carried into a hydrogen fueled flame which excites specific elements in the molecules, and the excited elements (P,S, Halogens, Some Metals) emit light of specific characteristic wavelengths.[5]The emitted light is filtered and detected by a photomultiplier tube.[5][4]In particular, phosphorus emission is around 510-536nm and sulfur emission os at 394nm.[5][4]
  • Atomic Emission Detector (AED), a sample eluting from a column enters a chamber which is energized by microwaves that induce a plasma.[5]The plasma causes the analyte sample to decompose and certain elements generate an atomic emission spectra.[5]The atomic emission spectra is defracted by a difraction gradient and detected by a series of photomultiplier tubes.[5]
  • Hall electrolytic conductivity detector (ElCD)
  • Helium ionization detector (HID)
  • Nitrogen-phosphorus detector (NPD),a form a thermionic detector where nitrogen and phosphorus alter the work function on a specially coated bead and a resulting current is measured.
  • Infrared detector (IRD)
  • Mass spectrometer (MS) - also called (GC-MS) highly effective and sensitive, even in a small quantity of sample.
  • Photo-ionization detector (PID)
  • Pulsed discharge ionization detector (PDD)
  • Thermionic ionization detector (TID)

Some gas chromatographs are connected to a mass spectrometer which acts as the detector. The combination is known as GC-MS. Some GC-MS are connected to an NMR spectrometer which acts as a backup detector. This combination is known as GC-MS-NMR. Some GC-MS-NMRare connected to an infrared spectrophotometer which acts as a backup detector. This combination is known as GC-MS-NMR-IR. It must, however, be stressed this is very rare as most analyses needed can be concluded via purely GC-MS.

[edit]MethodsThis image above shows the interior of a GeoStrata Technologies Eclipse Gas Chromatograph that runs continuously in three minute cycles. Two valves are used to switch the test gas into the sample loop. After filling the sample loop with test gas, the valves are switched again applying carrier gas pressure to the sample loop and forcing the sample through the Column for separation.

The method is the collection of conditions in which the GC operates for a given analysis.Method development is the process of determining what conditions are adequate and/or ideal for the analysis required.

Conditions which can be varied to accommodate a required analysis include inlet temperature, detector temperature, column temperature and temperature program, carrier gas and carrier gas flow rates, the column's stationary phase, diameter and length, inlet type and flow rates, sample size and injection technique. Depending on the detector(s) (see below) installed on the GC, there may be a number of detector conditions that can also be varied. Some GCs also include valves which can change the route of sample and carrier flow. The timing of the opening and closing of these valves can be important to method development.

[edit]Carrier gas selection and flow rates

Typical carrier gases include helium, nitrogen, argon, hydrogen and air. Which gas to use is usually determined by the detector being used, for example, a DID requires helium as the carrier gas. When analyzing gas samples, however, the carrier is sometimes selected based on the sample's matrix, for example, when analyzing a mixture in argon, an argon carrier is preferred, because the argon in the sample does not show up on the chromatogram. Safety and availability can also influence carrier selection, for example, hydrogen is flammable, and high-purity helium can be difficult to obtain in some areas of the world. (See: Helium-occurrence and production.) As a result of helium becoming more scarce, hydrogen is often being substituted for helium as a carrier gas in several applications.

The purity of the carrier gas is also frequently determined by the detector, though the level of sensitivity needed can also play a significant role. Typically, purities of 99.995% or higher are used. The most common purity grades required by modern instruments for the majority of sensitivities are 5.0 grades, or 99.999% pure meaning that there is a total of 10ppm of impurities in the carrier gas that could affect the results. The highest purity grades in common use are 6.0 grades, but the need for detection at very low levels in some forensic and environmental applications has driven the need for carrier gases at 7.0 grade purity and these are now commercially available. Trade names for typical purities include "Zero Grade," "Ultra-High Purity (UHP) Grade," "4.5 Grade" and "5.0 Grade."

The carrier gas linear velocity affects the analysis in the same way that temperature does (see above). The higher the linear velocity the faster the analysis, but the lower the separation between analytes. Selecting the linear velocity is therefore the same compromise between the level of separation and length of analysis as selecting the column temperature. The linear velocity will be implemented by means of the carrier gas flow rate, with regards to the inner diameter of the column.

With GCs made before the 1990s, carrier flow rate was controlled indirectly by controlling the carrier inlet pressure, or "column head pressure." The actual flow rate was measured at the outlet of the column or the detector with an electronic flow meter, or a bubble flow meter, and could be an involved, time consuming, and frustrating process. The pressure setting was not able to be varied during the run, and thus the flow was essentially constant during the analysis. The relation between flow rate and inlet pressure is calculated withPoiseuille's equation for compressible fluids.

Many modern GCs, however, electronically measure the flow rate, and electronically control the carrier gas pressure to set the flow rate. Consequently, carrier pressures and flow rates can be adjusted during the run, creating pressure/flow programs similar to temperature programs.

[edit]Stationary compound selection

The polarity of the solute is crucial for the choice of stationary compound, which in an optimal case would have a similar polarity as the solute. Common stationary phases in open tubular columns are cyanopropylphenyl dimethyl polysiloxane, carbowax polyethyleneglycol, biscyanopropyl cyanopropylphenyl polysiloxane and diphenyl dimethyl polysiloxane. For packed columns more options are available.[4]

[edit]Inlet types and flow rates

The choice of inlet type and injection technique depends on if the sample is in liquid, gas, adsorbed, or solid form, and on whether a solvent matrix is present that has to be vaporized. Dissolved samples can be introduced directly onto the column via a COC injector, if the conditions are well known; if a solvent matrix has to be vaporized and partially removed, a S/SL injector is used (most common injection technique); gaseous samples (e.g., air cylinders) are usually injected using a gas switching valve system; adsorbed samples (e.g., on adsorbent tubes) are introduced using either an external (on-line or off-line) desorption apparatus such as a purge-and-trap system, or are desorbed in the injector (SPME applications).

[edit]Sample size and injection technique[edit]Sample injection

The rule of ten in gas chromatography

The real chromatographic analysis starts with the introduction of the sample onto the column. The development of capillary gas chromatography resulted in many practical problems with the injection technique. The technique of on-column injection, often used with packed columns, is usually not possible with capillary columns. The injection system in the capillary gas chromatograph should fulfil the following two requirements:

  1. The amount injected should not overload the column.
  2. The width of the injected plug should be small compared to the spreading due to the chromatographic process. Failure to comply with this requirement will reduce the separation capability of the column. As a general rule, the volume injected, Vinj, and the volume of the detector cell, Vdet, should be about 1/10 of the volume occupied by the portion of sample containing the molecules of interest (analytes) when they exit the column.

Some general requirements which a good injection technique should fulfill are:

  • It should be possible to obtain the column's optimum separation efficiency.
  • It should allow accurate and reproducible injections of small amounts of representative samples.
  • It should induce no change in sample composition. It should not exhibit discrimination based on differences in boiling point, polarity, concentration or thermal/catalytic stability.
  • It should be applicable for trace analysis as well as for undiluted samples.
[edit]Column selection

The choice of column depends on the sample and the active measured. The main chemical attribute regarded when choosing a column is the polarity of the mixture, but functional groups can play a large part in column selection. The polarity of the sample must closely match the polarity of the column stationary phase to increase resolution and separation while reducing run time. The separation and run time also depends on the film thickness (of the stationary phase), the column diameter and the column length.

[edit]Column temperature and temperature program

A gas chromatography oven, open to show a capillary column

The column(s) in a GC are contained in an oven, the temperature of which is precisely controlled electronically. (When discussing the "temperature of the column," an analyst is technically referring to the temperature of the column oven. The distinction, however, is not important and will not subsequently be made in this article.)

The rate at which a sample passes through the column is directly proportional to the temperature of the column. The higher the column temperature, the faster the sample moves through the column. However, the faster a sample moves through the column, the less it interacts with the stationary phase, and the less the analytes are separated.

In general, the column temperature is selected to compromise between the length of the analysis and the level of separation.

A method which holds the column at the same temperature for the entire analysis is called "isothermal." Most methods, however, increase the column temperature during the analysis, the initial temperature, rate of temperature increase (the temperature "ramp") and final temperature is called the "temperature program."

A temperature program allows analytes that elute early in the analysis to separate adequately, while shortening the time it takes for late-eluting analytes to pass through the column.

[edit]Data reduction and analysis[edit]Qualitative analysis

Generally chromatographic data is presented as a graph of detector response (y-axis) against retention time (x-axis), which is called a chromatogram. This provides a spectrum of peaks for a sample representing the analytespresent in a sample eluting from the column at different times. Retention time can be used to identify analytes if the method conditions are constant. Also, the pattern of peaks will be constant for a sample under constant conditions and can identify complex mixtures of analytes. In most modern applications however the GC is connected to a mass spectrometer or similar detector that is capable of identifying the analytes represented by the peaks.

[edit]Quantitative analysis

The area under a peak is proportional to the amount of analyte present in the chromatogram. By calculating the area of the peak using the mathematical function of integration, the concentration of an analyte in the original sample can be determined. Concentration can be calculated using a calibration curve created by finding the response for a series of concentrations of analyte, or by determining therelative response factor of an analyte. The relative response factor is the expected ratio of an analyte to an internal standard (or external standard) and is calculated by finding the response of a known amount of analyte and a constant amount of internal standard (a chemical added to the sample at a constant concentration, with a distinct retention time to the analyte).

In most modern GC-MSsystems, computer software is used to draw and integrate peaks, and match MS spectra to library spectra.

[edit]Application

In general, substances that vaporize below ca. 300 °C (and therefore are stable up to that temperature) can be measured quantitatively. The samples are also required to be salt-free; they should not contain ions. Very minute amounts of a substance can be measured, but it is often required that the sample must be measured in comparison to a sample containing the pure, suspected substance known as areference standard.

Various temperature programs can be used to make the readings more meaningful; for example to differentiate between substances that behave similarly during the GC process.

Professionals working with GC analyze the content of a chemical product, for example in assuring the quality of products in the chemical industry; or measuring toxic substances in soil, air or water. GC is very accurate if used properly and can measure picomolesof a substance in a 1 ml liquid sample, or parts-per-billionconcentrations in gaseous samples.

In practical courses at colleges, students sometimes get acquainted to the GC by studying the contents of Lavender oil or measuring the ethylene that is secreted by Nicotiana benthamiana plants after artificially injuring their leaves. These GC analyses hydrocarbons (C2-C40+). In a typical experiment, a packed column is used to separate the light gases, which are then detected with a TCD. Thehydrocarbonsare separated using a capillary column and detected with an FID. A complication with light gas analyses that include H2is that He, which is the most common and most sensitive inert carrier (sensitivity is proportional to molecular mass) has an almost identical thermal conductivity to hydrogen (it is the difference in thermal conductivity between two separate filaments in a Wheatstone Bridge type arrangement that shows when a component has been eluted). For this reason, dual TCD instruments are used with a separate channel for hydrogen that uses nitrogen as a carrier are common. Argon is often used when analysing gas phase chemistry reactions such as F-T synthesis so that a single carrier gas can be used rather than 2 separate ones. The sensitivity is less but this is a tradeoff for simplicity in the gas supply.

[edit]GCs in popular culture

Movies, books and TV shows tend to misrepresent the capabilities of gas chromatography and the work done with these instruments.

In the U.S. TV show CSI, for example, GCs are used to rapidly identify unknown samples. For example, an analyst may say fifteen minutes after receiving the sample: "This is gasolinebought at a Chevronstation in the past two weeks."

In fact, a typical GC analysis takes much more time; sometimes a single sample must be run more than an hour according to the chosen program; and even more time is needed to "heat out" the column so it is free from the first sample and can be used for the next. Equally, several runs are needed to confirm the results of a study - a GC analysis of a single sample may simply yield a result per chance (see statistical significance).

Also, GC does not positively identify most samples; and not all substances in a sample will necessarily be detected. All a GC truly tells you is at which relative time a component eluted from the column and that the detector was sensitive to it. To make results meaningful, analysts need to know which components at which concentrations are to be expected; and even then a small amount of a substance can hide itself behind a substance having both a higher concentration and the same relative elution time. Last but not least it is often needed to check the results of the sample against a GC analysis of a reference sample containing only the suspected substance.

A GC-MS can remove much of this ambiguity, since the mass spectrometer will identify the component's molecular weight. But this still takes time and skill to do properly.

Similarly, most GC analyses are not push-buttonoperations. You cannot simply drop a sample vial into an auto-sampler's tray, push a button and have a computer tell you everything you need to know about the sample. The operating program must be carefully chosen according to the expected sample composition.

A push-button operation can exist for running similar samples repeatedly, such as in a chemical production environment or for comparing 20 samples from the same experiment to calculate the mean content of the same substance. However, for the kind of investigative work portrayed in books, movies and TV shows this is clearly not the case.

What is a meniscus in chemistry?

In chemistry, the meniscus refers to the curve seen at the surface of a liquid inside a container. It is caused by the surface tension of the liquid, and is used to accurately measure the volume of a liquid in a graduated cylinder or similar vessel.

What is Chromatography and what is it for?

Chromatography is an analytical method to separate the compounds from the two immiscible phases,one being a mobile phase and another being a stationary phase. It is used in petroleum industry,pharmaceutical industry.

What does chromatography mean in Greek?

The name Chromatography comes from the greek word "Chromos" meaning color and "grafein" meaning to write. So, Chromatography is "writing with colors"...

What is an eluent?

An eluent is a solvent or liquid used in chromatography to carry the sample through the stationary phase, enabling the separation of the components based on their chemical properties. It is important for the eluent to be compatible with the stationary phase and the sample being analyzed to achieve efficient separation.

Gas chromatography for the detection of hydrogen?

Gas chromatography is a technique used to separate and analyze components in a gas sample. It works by passing the gas sample through a column where the different components are separated based on their interactions with the column material. Detection of hydrogen in gas chromatography is typically done using a thermal conductivity detector, which is sensitive to changes in thermal conductivity caused by the presence of different gas components.

Principle of partition chromatography?

Chromatography in which separation is based mainly on differences be-

tween the solubility of the sample components in the stationary phase (gas

chromatography), or on differences between the solubilities of the compo-

nents in the mobile and stationary phases (liquid chromatography)

What can chromatography be used for?

The uses of chromatography are important in checking the purity of oils extracted from plants, which are often used in the medicine industry. It shows all the different colours of chemicals in the mixture can be used to compare the elements of two mixtures.

What is paper chromatography?

Paper chromatography is a laboratory technique used to separate and analyze mixtures of substances. It involves applying a sample to a strip of paper, which is then placed in a solvent and allowed to separate into its individual components based on their different affinities to the paper and the solvent. The separated components can be visualized as distinct bands on the paper strip.

What are the different types of chromatography?

there are different types like gas chrom. and thinlayer chrom Answer: There are two types of chromatography:liquid chromatography gas chromatography

What happens to water molecules when they are heated?

Like all molecules, a molecule of H20 is in constant motion; 'hot' molecules move faster than 'cold' molecules. If the molecules move slowly enough the substance appears stationary to us (frozen water) and if they move quickly enough they will expand to fill their container (vapor or gas water.)

Inflation is undesirable because it?

Inflation is considered to be undesirable because it arbitrarily redistributes wealth and real income. It also causes consumers to pay more for goods and services and causes the value of the dollar to go down.

Disadvantages of paper chromatography?

  1. The only disadvantage, like any paper chromatography method, is that the test take a lot of time - typically 45 to 90 minutes.
  2. This method also yield little amount of pigments when it comes to the extraction of the isolated pigments. Several tests must be run to gather a respectable amount for further chemical tests.

What might be some limitations to paper Chromatography?

Some limitations of paper chromatography include limited resolution due to overlapping components, potential separation difficulties for compounds with similar properties, and the risk of sample contamination if the paper is not handled properly. Additionally, the process can be time-consuming and may require optimization for specific compounds or analytes.

What is the significance of TLC over paper chromatography?

Paper chromatography and TLC both use a solid stationary phase and liquid mobile phase. The only real difference is in the physical and chemical composition of the stationary phase. Paper chromatography uses a fibrous protein/polymer mesh (i.e. paper) as its stationary phase, TLC uses a compacted powder (usually silica or alumina) on a sealed backing.

Why is filter paper used in chromatography?

Filter paper is used in chromatography because it is porous and allows the solvent to travel through it by capillary action. This helps separate the components of a mixture based on their solubility and affinity for the stationary phase. The filter paper also provides a large surface area for the separation to occur effectively.

What is the difference between partition chromatography and adsorption chromatography?

The partition chromatography involves separation between liquids while adsoption chromatography involves solid and liquid separations. Answer: Partition Chromatography is a liquid liquid extraction which involves two solvents while adsorption chromatography is a liquid solid extraction which involves a solid stationary phase & a liquid mobile phase.

Example for writing of test scenario?

EX:Login page In login page you have to give values for user name, password and then click on ok button to login and cancel button to close the login window user name:Alphanumerics with 4-16characters long password:lowercase letters with 4-8 chars long OK :Next window CANCEL :Close the window Prepare Test Scenarios Test Scenario Template --------------------------- Test scenario1:Verify user name value Test Scenario2:verify pass word value Test Scenario3:verify OK button operation to login Test Scenario4:verify cancel button operation to close the window Ts1:Verify user name value a)Boundary value analysis (on Size) min=4characters maximum=16 characters more than max or less than min are not allowed b)Equivalence class partitions(on type) A-Z or a-z or 0-9 are valid types to be entered special symbols or blank fields are in valid Test Scenario2:verify pass word value SAME AS USER NAME Test Scenario3:verify OK button operation to login Decision table user name ------pass word-----expected outcome after click on "OK" valid valid next window valid invalid error message invalid valid error message blank value error message value blank error message Test Scenario4:verify cancel button operation to close the Decision table user name ------pass word-----expected outcome blank blank close the window value blank close the window blank value close the window value value close the window Like the above example we will write test scenarios for all the Functional Specifications.